1. Core principle: "micro-contour capture" of stylus + inductive sensor
The core principle of the stylus roughness meter (also known as the stylus tracer) is to convert mechanical displacement into electrical signals through contact measurement between the diamond stylus and the measured surface:
Stylus sliding: The stylus (the curvature radius of the tip is extremely small) is vertically placed on the surface of the workpiece and slides at a constant speed. The peaks and valleys of the surface will drive the stylus to move up and down;
Signal conversion: The displacement of the stylus is transmitted to the magnetic core through the fulcrum, causing the inductance of the surrounding differential inductor coil to change, which in turn causes the balance bridge to be unbalanced and output an electrical signal proportional to the displacement;
Signal processing: After amplification and phase-sensitive detection, the weak electrical signal is divided into three outputs - the zero indicator displays the stylus position, the recorder draws the contour graph, and the integral calculator calculates the surface roughness Ra value (arithmetic mean deviation).
2. Measurement process: key steps from zeroing to reading
Zeroing calibration: ensure that the core is in the middle of the inductor coil, the pointer of the zeroing meter is zeroed, and the measurement is within the linear range of inductance change;
Filtering: remove irrelevant signals through noise filtering and waviness filtering, and retain valid signals related to roughness;
Parameter calculation: after the average meter amplifier, the signal current is proportional to the absolute value of the height of the surface profile deviating from the center line. After the integrator completes the calculation, the Ra value is directly displayed (some models can measure parameters such as Rz, Ry).
Scope of application: can measure complex-shaped workpieces such as planes, cylinders, cones, small holes, grooves, etc., the Ra value measurement range is 0.02-10μm (some models can measure smaller values), high accuracy and convenient operation.
3. Limitations of traditional models: These shortcomings need attention
Although stylus roughness testers are widely used, traditional models have obvious shortcomings:
Single parameters: only a few parameters such as Ra, Rz, and Ry can be measured, which is difficult to meet the comprehensive evaluation of complex surfaces;

Limited range: The upper limit of Ra value measurement is mostly 10μm, which cannot cover ultra-rough or ultra-smooth surfaces;
Low flexibility: The adjustment of parameters such as evaluation length and filter is limited, and the adaptation scenario is narrow;
Output is not intuitive: The analog signal processing error is large, and the results are mostly numerical display, lacking graphical analysis;
Poor system reliability: It is greatly affected by environmental vibration and stylus wear, and the long-term stability is insufficient.
The core reason for these problems is that traditional models rely on analog signal processing, errors are easy to accumulate and difficult to optimize digitally.
4. Summary: Applicable scenarios of stylus roughness testers
It is suitable for medium and high precision Ra value measurement (0.02-10μm) of hard surfaces such as metals and ceramics, especially for planes, shafts, small holes and other workpieces in mechanical processing. However, please note that if multi-parameter, large-scale or digital analysis is required, it is recommended to choose an upgraded smart model.